Met One Aerocet-532 Mass Particle Counter/Dust Monitor
2-in-1 performance! Particle Mass Profiler and Counter in a Single Handheld Unit Six Mass Ranges and Four Particle Sizes Tailored Mass Conversion Flexible Data Interfaces Temperature and Relative Humidity Option |
|
Aerocet 531S
|
Specifications
OPERATING PRINCIPLE | Counts individual particles using scattered laser light and calculates the equivalent mass concentration using a proprietary algorithm |
PERFORMANCE | |
Mass Mode | |
Mass Concentration Ranges | PM1, PM2.5, PM4, PM7, PM10, and TSP |
Concentration Range | 0 – 1000 ug/m3 |
Sample Time – Mass Mode | 1 minute |
Particle Mode | |
Particle Size Range | Four channels—0.5, 1.0, 5.0 and 10.0 μm |
Concentration Range | 0 – 3,000,000 particles per cubic foot (105,900 particles/L) |
Sample Time – Particle Mode |
1 minute |
Accuracy | ± 10%, to calibration aerosol |
Sensitivity | 0.5 μm |
Flow Rate | 0.1 cfm (2.83 lpm) |
ELECTRICAL | |
Light Source | Laser diode, 90 mW, 780nm |
Power | 7.4V Li-ion battery pack—provides 20 hours of typical intermittent operation, up to 10 hours continuous use |
AC Adapter/Charger | L-ion battery charger, 100-240 VAC, 50/60Hz, 0.2A |
Communications | RS-232, USB |
Particle Certifications | Meets or exceeds CE and ISO21501 |
Mass Certifications | Requires proper use of appropriate K-factors for material being measured |
INTERFACE | |
Display | 16-character x 4-line LCD |
Keyboard | 7-key membrane type |
PHYSICAL | |
Size | Height = 6.25" (15.9 cm) Width = 4.0" (10.2 cm) Thickness = 2.1" (5.4 cm) |
Weight | 2 lb – 32 oz (0.9 kg) |
ENVIRONMENTAL | |
Operating Temperature | 0° to +50°C |
Storage Temperature | -20° to +60°C |
ACCESSORIES | |
Supplied | Operation Manual Custom Serial Cable USB Cable Communication Software (Comet) AC Adapter / Battery Charger Isokinetic Sample Probe Carrying Case Zero Particulate Filter |