PM10 High Volume Air Sampler (VFC, Digital Timer)

SKU 6070DV
Original price $8,457.50 - Original price $8,457.50
Original price
$8,457.50
$8,457.50 - $8,457.50
Current price $8,457.50

PM10 is the designation for particulate matter in the atmosphere that has an aerodynamic diameter of 10 micrometers (μm) or less.

A high volume PM10 sampler draws a known volume of ambient air at a constant flow rate through a size selective inlet and a filter. Particles in the PM10 size range are then collected on the filter during the specified 24-hour sampling period. Each sample filter is weighed before and after sampling to determine the net weight (mass) gain of the collected PM10 sample.

The total volume of air sampled is determined from the measured volumetric flow rate and the sampling time. The concentration of PM10 in the ambient air is computed as the total mass of collected particles in the PM10 size range divided by the volume of air sampled. This sampled air must be corrected to U.S. EPA standard conditions (25°C, 760mm Hg or 101 kPa), and the PM10 measurement is expressed as micrograms per standard cubic meter (μg/std. m3).

Design

A high volume PM10 sampler consists of two basic components: a specially designed inlet that transmits only those particles less than 10μm in diameter and a flow control system capable of maintaining a constant flow rate within the design specifications of the inlet.

The symmetrical design of the PM10 impaction inlet insures wind-direction insensitivity. Ambient air is drawn into the inlet and is evacuated from a buffer chamber through nine acceleration nozzles. By virtue of their larger momentum, particles greater than 10 microns aerodynamic diameter impact onto a greased impaction shim plate. The particles smaller than 10 microns are carried vertically upward by the air flow and down sixteen vent tubes to the standard 8 x 10-inch (20.3 x 25.4cm) filter where they are collected. The acceleration nozzles have critical diameters to provide the necessary changes in velocity to effect correct particle size fractionation within the impaction chamber.

Two types of flow control systems are available. One is equipped with a mass flow control system; the other is equipped with a volumetric flow control system.

The flow rate in a mass flow controlled system is actively sensed and controlled at some predetermined set point. Air is pulled through the filter into the intake of a blower and subsequently exits the sampler through an exit orifice, which facilitates measurement of the flow with a manometer or pressure recorder. The flow rate is controlled by an electronic mass flow controller which uses a flow sensor probe installed below the filter holder to monitor the mass flow rate and to control the speed of the motor accordingly. The controlled flow rate can be changed by an adjustment knob on the flow controller.

A volumetric flow control system maintains a constant volumetric flow rate (given a fixed temperature) through the PM10 inlet. A choked-flow venturi is operated such that the air attains sonic velocity in the throat of the device. In this "choked" mode, the flow rate is unaffected by downstrem conditions such as motor speed or exit pressure, and is a predictable function of upstream conditions such as the stagnation pressure ratio and temperature. The volumetric flow is controlled without any moving parts or electronic components.

Features

Accurately samples particles with aerodynamic diameter less than 10 micrometers (PM10)

Specifications

Flow Rate 40 cfm (1.13 m3/min)